Sisco Care

Instruments Cleaning & Care

SiscoCare Guidelines; Process/Material Requirements:

Water PH, Chloride & Fluoride
  • pH: 4.5 - 9.5
  • Chloride: Less than 120 mg/l (ppm)
  • Fluoride: Less than 200 mg/l (ppm)
  • Water Temperature: Below 45°C
Intended Use
  • Instruments are meant for their specific intended use only.
  • Allow instruments to reach room temperature before opening after autoclave or operation.
Post-operation Cleaning Start Time:
  • Start cleaning within 6 hours (recommended 20 minutes).
  • Use approved wetting agents.
  • Ensure racks, blades, and working parts are in the open position for better exposure.
During Cleaning
  • Use separate racks for delicate, small, micro-structured, and heavy instruments.
  • Avoid dropping or bumping instruments.
  • Use genuine cleaners and disinfectants, free from foam carryover and residues.
  • Immerse instruments for approximately 5 minutes in hydrogen peroxide.
  • Use demineralized or distilled water, free from minerals, silicates, and chlorides for final rinsing.
Soap Solution
  • Use freshly prepared water daily.
  • Use neutral pH soap (between 7 - 8), free from iodine, chlorhexidine, and bleaches.
Cleaning
  • Clean only stainless steel instruments.
  • Use stiff nylon brushes under water.
Drying
  • Wipe instruments with soap-free towels and air dry.
  • Maintain room temperature and constant humidity.
Storage
  • Use separate racks for delicate, small, micro-structured, and heavy instruments.
  • Store instruments properly after drying.
  • Ensure storage is free from stress and kinks, and free from corrosive fumes.
Sterilisation
  • Autoclave at 132°C for 15 minutes (15 - 20 psi) using distilled water.
  • Ensure adequate drying after autoclaving.
  • Autoclave instruments with ratchets in the open position.
  • Place heavy instruments at the bottom and delicate ones at the top.
Lubrication
  • Use paraffin or white oil for lubrication.
  • Follow manufacturer’s instructions for exposure time and concentration during sterilization soaking in disinfectant solution.
  • Perform final rinse with sterile distilled water and ensure proper drying.